The dosimetric impact of setup errors in image guided radiotherapy
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
University of the Witwatersrand, Johannesburg
Abstract
Aim: This study aimed to contribute to clinical implementation of imaging protocols in Image-Guide Radiotherapy (IGRT) by evaluating the impact of positional shifts on the dose distribution and target coverage in Volumetric Modulated Arc Therapy (VMAT) for nasopharynx cancer. The study focused on understanding the dosimetric impact of these shifts on Planning Target Volumes (PTVs) and Organs at Risks (OARs). In radiotherapy, the adoption of advanced treatment techniques made possible by technological advancements demands higher levels of setup accuracy, necessitating standard operating procedures in IGRT and associated quality assurance (QA) program that enhance patient plan verification and treatment accuracy without compromising treatment efficacy. Methodology and Materials: The measurement were performed on anthropomorphic CIRS Shoulder, Head and Neck (SHANE) phantom, simulating patient anatomy to assess the effect of positional shifts in VMAT plans for nasopharyngeal cancer. In accordance with IAEA and QUANTEC guidelines, a VMAT plan with at least two full rotation beams (arcs) was created with a dose prescription of 70 Gy distributed in 35 fractions, giving 2.0 Gy per fraction to PTV_7000. Corresponding copied plans from the original VMAT were developed by introducing positional shifts to the isocenter in all directions. These copied plans were recalculated and evaluated in the Elekta Monaco 6.1.2.0 treatment planning system (TPS) using plan Dose-Volume Histogram (DVH) data to analyze the dosimetric impact. For Patient Specific Quality Assurance (PSQA), the Octavius 4D phantom with 1500-chamber array was employed to assess the impact of these positional shifts on dose distribution using VeriSoft software (version 7.2). The measured dose distribution resulting from these translational shifts was compared with the planned dose distribution from TPS using Gamma analysis method. Data from the DVH analyses and gamma passing rates were employed to determine shift tolerances, which were then compared to literature-based tolerances. Results: The study investigated the impact of positional shifts on dose distribution in VMAT for nasopharyngeal cancer, focusing on maintaining optimal PTV coverage and sparing OARs. Results showed that translational shifts greater than ±2 mm reduces the PTV dose coverage, with the D % > 62.7 Gy criterion failing beyond ±2 mm and decreasing by up to 50% of the initial PTV dose coverage at maximum shifts (±20 mm), potentially compromising treatment efficacy. DVH statistics showed that longitudinal and vertical shifts, particularly with inferior and posterior shifts, resulted in decrease of PTV dose coverage by up to 70% and 40%. Hot spots increase with lateral, longitudinal, and vertical shifts, posing the risks of unintended dose escalation. Dose distribution to OARs, including spinal cord, brainstem, brain, and parotid glands, varied depending on positional shifts, increasing risk of overdosing these critical organs. PSQA further highlighted the importance of accurate and precise patient positioning, with lateral and vertical shifts resulting in gamma passing rate failure, and established a shift tolerances of 4-6 mm. The results highlight the critical need for imaging protocols to account for setup errors to ensure target conformity, optimal dose delivery, and treatment efficacy. Conclusion: The dosimetric impact of positional shifts in nasopharynx patients undergoing VMAT radiotherapy and Shane phantom were investigated. The study showed that even minor shifts effect the dose coverage of the OARs and PTV, potentially compromising patient safety and treatment efficacy. The study emphasized the importance of accurate patient positioning, strict adherence to the established shift tolerances (±2 mm), and verification techniques such as daily or weekly IGRT (e.g., CBCT), PSQA, and adaptive planning. Overall, the study emphasize a rigorous patient position verification process and regular use of QA tools like the Octavius® 4D system, as well as adherence to established shift tolerances to maintain effective dose delivery, particularly in sensitive regions like nasopharynx, where proximity of critical organs requires accurate and precise treatment planning.
Description
A research report submitted in fulfilment of the requirements for the degree of Master of Science (Medical Physics) to the Faculty of Science, School of Physics, University of the Witwatersrand, Johannesburg, 2025
Citation
Maabane, Berry. (2025). The dosimetric impact of setup errors in image guided radiotherapy. [Master's dissertation, University of the Witwatersrand, Johannesburg]. WIReDSpace. https://hdl.handle.net/10539/50319