Beam quality specification in kilo-voltage radiotherapy

DSpace/Manakin Repository

Show simple item record

dc.contributor.author Jozela, Sibusiso
dc.date.accessioned 2008-05-29T13:00:02Z
dc.date.available 2008-05-29T13:00:02Z
dc.date.issued 2008-05-29T13:00:02Z
dc.identifier.uri http://hdl.handle.net/10539/4902
dc.description.abstract Purpose: The purpose of this work was to compare and analyse two clinically measurable beam quality specifiers, the half value layer (HVL) and the ratio of the doses at depths 2 cm and 5 cm (D2/D5) for a range of kilovoltage modalities, and to determine whether a practical, alternative and/or better correlation exists. Methods and materials: Four x-ray units were used: two Philips RT 250 units, a Pantak HF 420 operated up to 250 kV, and a D3300 Gulmay Medical unit operated up to 300 kV. As not all these units were equipped with an internal monitor chamber, a system was used where either the first measurement was repeated at the end of each series or an external monitor chamber was employed in order to ensure output constancy. A range of HVL’s were measured on each of the energies investigated on this work, which were used clinically. A calibrated 0.6 cc ionization chamber was used in a 30 cm x 30 cm x 30 cm water phantom to measure the absorbed dose to water at depths 2 cm and 5 cm in order to investigate D2/D5 as the alternative quality index. Results: The effectiveness of using a monitor chamber in the determination of HVL has been shown to be significant in this work where HVLs differed by up to 3%. Errors incurred from using HVL have been identified. This work verified that the ratio of doses at depths 2 cm and 5 cm in water could be applied as a kilovoltage beam quality specifier in the clinical environment at low and medium energies with a well defined FSD and field size. Conclusions: The use of D2/D5 as a tool to verify the beam quality index would simplify quality control in the clinical environment. Further work would have to be done to investigate other energies. Lower energies may require the use of shallower depths in order to improve accuracy and ensure a more clinically relevant setup. en
dc.format.extent 2567374 bytes
dc.format.extent 16404 bytes
dc.format.mimetype application/pdf
dc.format.mimetype application/pdf
dc.language.iso en en
dc.subject HVL en
dc.subject D2/D5 en
dc.subject dosimetry en
dc.title Beam quality specification in kilo-voltage radiotherapy en
dc.type Thesis en


Files in this item

This item appears in the following Collection(s)

Show simple item record

Search WIReDSpace


Advanced Search

Browse

My Account

Statistics